Chromel / Alumel · ITS-90

Type K Thermocouple Calculator

EMF ↔ temperature with cold-junction compensation, plus IEC 60584-1 / ASTM E230 limits-of-error lookup. Math from the NIST ITS-90 reference functions.

Thermometrics Corp →
Enter a voltage read across the thermocouple leads and the reference-junction temperature.
Valid −5.891 to 54.886 mV
Meter / terminal-block temperature. Use 0 for an ice-point reference.
Hot-junction temperature
—

Request a quote with these values →

Passes the computed configuration to the Thermometrics RFQ form — no re-entry.

A Type K thermocouple is a base-metal sensor with a chromel (~90% Ni, 10% Cr) positive leg and an alumel (~95% Ni, with Al, Mn, Si) negative leg. It spans roughly −270 °C to +1372 °C on the ITS-90 scale with a sensitivity near 41 µV/°C, which makes it the default general-purpose thermocouple in industrial service.

What standards govern a Type K thermocouple?

Two standards apply. The EMF–temperature relationship is fixed by the ITS-90 reference function in NIST Monograph 175, and the permissible error is set either by IEC 60584-1 (Class 1, 2, 3) or by ASTM E230 / ANSI MC96.1 (Standard and Special limits of error).

This site evaluates that reference function directly: a piecewise polynomial for the −270 °C to 0 °C and 0 °C to 1372 °C ranges, with the additional exponential term that Type K requires above 0 °C. Conversion back from voltage uses the NIST inverse polynomials, whose residual fit error is roughly −0.05 to +0.06 °C across the working range — small relative to the sensor's own limits of error. The two tolerance regimes are not interchangeable; the tolerance class page sets them side by side. For the full chromel/alumel construction range, see the Type K thermocouple catalog.

Polarity and color code are a frequent source of field error: under ANSI MC96.1 the Type K overall jacket and positive (KP) leg are yellow and the negative (KN) leg is red; under IEC 60584-3 the jacket and positive leg are green and the negative leg is white. A reversed leg produces a plausible-but-wrong reading rather than an open circuit.

How do you convert a Type K millivolt reading to temperature?

Measure the EMF across the leads and the reference-junction temperature, add the reference junction's table EMF to the reading, then invert the ITS-90 curve. The converter above does exactly this. Compensating in temperature instead of voltage introduces error because the response is non-linear.

E_total = E_measured + E_ref( T_cold_junction )
T_hot = InverseITS90( E_total )

A thermocouple generates an EMF proportional to the temperature difference between its hot and reference junctions — the Seebeck effect — not absolute temperature. For manual lookups the printable millivolt chart carries the same ITS-90 values at 1 °C resolution. For a primer on signal conditioning and CJC hardware, see cold-junction compensation notes.

Why do Type K readings go wrong?

The characteristic Type K degradation modes are preferential oxidation (“green rot”) in low-oxygen atmospheres near 800–1050 °C, the ~354 °C alumel magnetic anomaly, EMF aging and calibration drift, insulation-resistance shunting from moisture, and mechanical or vibration fatigue at the junction.

Each mode has a distinct symptom signature, which makes them diagnosable in the field. The troubleshooting guide works through them symptom-first; installation practice prevents most of them. Recalibration and certificate services are covered under thermocouple calibration.

Frequently asked questions

What temperature range does a Type K thermocouple cover?
A Type K thermocouple covers approximately −270 °C to +1372 °C on the ITS-90 scale. The usable continuous range is narrower and depends on wire gauge, sheath, and atmosphere. Sensitivity is about 41 µV/°C near room temperature.
How accurate is a Type K thermocouple?
Accuracy on a new sensor is set by its tolerance class. IEC 60584-1 Class 1 is the greater of ±1.5 °C or 0.4% of the temperature; Class 2 is ±2.5 °C or 0.75%. ASTM E230 Special is ±1.1 °C or 0.4%; Standard is ±2.2 °C or 0.75%. In-service drift is additional.
What is cold-junction compensation and why is it required?
A thermocouple measures the EMF produced by the temperature difference between its hot and reference junctions, not absolute temperature. Because the reference junction is rarely at 0 °C, you must add its table EMF back before inverting the curve. Compensating in temperature instead of voltage adds error because the response is non-linear.
What colour is Type K thermocouple wire?
Under ANSI MC96.1 the Type K jacket and positive (chromel) leg are yellow and the negative (alumel) leg is red. Under IEC 60584-3 the jacket and positive leg are green and the negative leg is white. A reversed leg produces a wrong reading, not an open circuit.
Why does a Type K thermocouple drift or read low?
Common causes are preferential oxidation (green rot) in low-oxygen atmospheres between about 800 °C and 1050 °C, cold-junction compensation error, EMF aging from cumulative time at temperature, and insulation-resistance shunting from moisture at the termination.
What are chromel and alumel?
Chromel is the positive leg, roughly 90% nickel and 10% chromium. Alumel is the negative leg, roughly 95% nickel with aluminium, manganese, and silicon. The pairing defines the Type K thermoelectric response.